InfraLUM FT-02 Fourier-Transform Spectrometer |
|
InfraLUM FT-02 is a multipurpose mid-IR FT spectrometer designed for laboratory use. The instrument is equipped with a data acquisition and processing system operating with a dedicated PC SpectraLUM analytical software package.
Description
|
|
- Full-size sample compartment
- LUMEX proprietary DCE interferometer
- Permanently aligned optics
- Sealed and desiccated optics
- Shortened optical beam path
- User-friendly software
- Advanced data manipulation
- Standard accessories applicable
- Built-in diagnostics
|
|
A proprietary "double cat's eye" design of the interferometer, which is vibration- and misalignment-proof, ensures high measurement stability, accuracy and precision. The Fourier transform of the detected interferogram reconstructs the IR spectrum throughout the spectral range. Fast interferogram detection (0.3 - 2.5 s, depending on resolution) allows multiple measurements to be made during a short period of time, which improves the signal-to-noise ratio. The large sample compartment can accommodate all conventional and non-standard large cells and accessories. The data acquisition and processing system, operating under Windows® 95/98 operation system, provides simple control and allows even unskilled personnel to easily master the instrument operation.
| PERFORMANCE CHARACTERISTICS AND SPECIFICATIONS
|
|
Spectral range - total |
6500 - 350 cm-1 |
|
Spectral range - with validated metrological characteristics |
4500 - 550 cm-1 |
|
Resolution |
0,5, 1, 2, 4, 8, and 16 cm-1 |
|
Spectrum measuring time |
60 sec |
|
Power requirements |
110/220 Vac, 50/60 Hz |
|
Power consumption |
60 W |
|
Dimensions |
580x515x295 mm |
|
Weight |
37 kg |
|
Wavenumber accuracy |
0.05 cm-1 |
|
Photometric accuracy |
< 0.1% T |
|
Nonuniformity of a 100-% line |
< 0.4% T |
|
Radiation source |
ceramic |
|
Signal-to-noise ratio* |
8000 |
|
Detector |
pyroelectric |
|
Beamsplitter |
Ge/KBr (Ge/ZnSe) |
|
Power requirements |
110/220 V, 60 W |
|
* For an RMS deviation of a 100-% line in the 2000-2200 cm-1 spectral region with a resolution of 4 cm-1, the measurement time is 1 min.
For additional information on features and applications of the system, please refer to our presentation: "mid-IR FT spectrometry with InfraLUM FT-02. Àpplications."
(Download mid-IR-applications.zip 203 Kb).
|